AC HI-POT
|
Voltage Range |
0.100 ~ 5.000kV |
Voltage Resolution |
5V |
Voltage Regulation |
1% + 5 V |
Voltage Accuracy |
¡¾(1% of reading ¡¾10 V) |
Max. Current Sourcing |
30mA ~ 40mA (above 500V, maximum test time 180sec) 0.1mA ~ 29.99mA (above 500V, continuous test) 0.1mA ~ 40mA (below 500V, continuous test) |
Current Limit |
0.10 ~ 40mA, 0.02mA/step |
Current Accuracy |
¡¾(1% of reading ¡¾80¥ìA) |
DC HI-POT
|
Voltage Range |
0.100kV ~ 6.000kV |
Voltage Resolution |
5V |
Voltage Regulation |
1% + 5 V |
Voltage Accuracy |
¡¾(1% of reading ¡¾10 V) (above 500V) |
Max. Current Sourcing |
0.1mA ~ 10mA (above 500V, continous test) 0.1mA ~ 2mA (below 500V, continuous test) |
Current Limit |
0.1mA ~ 10mA, 0.01mA/step |
Current Accuracy |
¡¾(1% of reading ¡¾50¥ìA) |
INSULATION RESISTANCE
|
DC Voltage |
50V/100V/500V/1000V |
Resistance Range |
1 ~ 9900M¥Ø(500V,1000V) 1 ~ 2000M¥Ø(50V,100V) |
Resistance Accuracy |
50V/100V 1M¥Ø ~ 50M¥Ø: ¡¾(5% of reading ¡¾1count) 51M¥Ø ~ 200M¥Ø: ¡¾(10% of reading ¡¾1count) 201M¥Ø ~ 1990M¥Ø: ¡¾(20% of reading ¡¾1count) |
|
500V/1000V 1M¥Ø ~ 500M¥Ø: ¡¾(5% of reading ¡¾1count) 501M¥Ø ~ 2000M¥Ø: ¡¾(10% of reading ¡¾1count) 2001M¥Ø ~ 9900M¥Ø: ¡¾(20% of reading ¡¾1count) |
CONTINUITY CHECK
|
Output Current Range |
0.1A (DC) |
Compliance Voltage Range |
10% |
Resolution |
0.1¥Ø @ 1¥Ø |
RAMP TIME & TEST TIME
|
AC Hi-Pot Ramp/Test Time |
0.0 ~ 999.9 seconds |
DC Hi-Pot Ramp/Test Time |
0.0 ~ 999.9 seconds |
Insulation Resistance Ramp/Test Time |
0.0 ~ 999.9 seconds |
Continuity Check Ramp/Test Time |
0.0 ~ 999.9 seconds |
ARC DETECT
|
Detect Current |
40 level (1 ~ 40mA) |
INTERFACE
|
Remote I/O |
Standard |
RS-232C |
Standard |
GPIB |
Option |
Scanner Box |
Option (HSB-001-1 : 8ch HV+8ch GB ; HSB-001-2 : 16ch HV) |
POWER SOURCE |
|
|
AC 100V/120V/220V/230V ¡¾ 10%, 50/60Hz |
ACCESSORIES
|
|
User manual x 1 Power cord x 1 Test lead GHT-105A x 1 |
DIMENSIONS & WEIGHT |
|
|
330(W) x 149(H) x 446(D) mm Approx.15kg |