Insulation Resistance |
Test Voltage |
250V/500V/1000V |
Measuring Ranges (Mid-scale Value) |
100M¥Ø/200M¥Ø/400M¥Ø (1M¥Ø) (2M¥Ø) (4M¥Ø) |
Output Voltage on open circuit |
Rated test voltage+20%, -0% |
Nominal Current |
1mA DC min. |
Output Short Circuit Current |
1.3 mA DC approx. |
Accuracy |
0.1~10M¥Ø/0.2~20M¥Ø/0.4~40M¥Ø (Accuracy Guaranteed Ranges) ¡¾5% of indicated value |
Continuity |
Measuring Ranges (Mid-scale Value) |
2¥Ø/20¥Ø (1¥Ø)(10¥Ø) |
Output Voltage on open circuit |
4~9V DC |
Measuring Current |
200mA DC min. |
Accuracy |
|¡¾3% of scale length |
General |
Safety Standard |
IEC61010-1 CAT.III 300V Pollution Degree 2 IEC61010-2-031 IEC61557-1/2/4 |
IP Rating |
IEC60529-IP54 |
EMC Standard |
IEC61326-1 |
Withstand Voltage |
3700V AC for 1 minute |
Power Source |
R6P(AA)(1.5V) ¡¿ 6 |
Dimensions |
185(L) ¡¿ 167(W) ¡¿ 89(D)mm |
Weight |
860g approx. |
Accessories |
7122(Test leads) Pouch for test leads 8923(F500mA/600V fuse) ¡¿ 2 R6P(AA) ¡¿ 6 Shoulder Strap Instruction Manual |